The Kowa FM-700 is an aqueous flare meter with a slit-lamp design. Even in cases of corneal inflammation, anyone with slit lamp experience will find the alignment process intuitively easy. It is also applicable for monitoring patients post cataract surgery and for testing ocular inflammatory diseases such as uveitis.
An area including the measurement window is scanned with a laser beam. Background signal 1 (BG1) is obtained when the laser beam is located below the measurement window and background signal 2 (BG2) is obtained when the laserbeam is located above the measurement window. Both are scattered light noise from intraocular tissue, whilst the flare signal (SIG) is a sum of scattered light from protein and scattered light noise from intraocular tissue.
Thus, the intensity of the scattered light caused by the protein concentration in the aqueous humor of the anterior chamber is calculated using the formula: SIG-(BG1+BG2)/2. The result obtained from using this formula is called ‘flare value Eand represented as photon count per millisecond.
Binokulares stereoskopisches Mikroskop Binocular stereoscopic microscope | Vergrößerung 7,10,16,26,40x Magnification 7,10,16,26,40x |
Dioptrien Ausgleichs-Bereich des Okulares Dioptres Compensation range of the eyepiece | ±5D ±5D |
Augenabstandseinstellung eye-to-eye distance adjustment | 55 – 72mm 55 – 72mm |
Spaltbreite slit width | 0 – 11mm kontinuierlich 0 - 11mm continuous |
Spaltlänge slit length | 1 – 9mm kontinuierlich 1 - 9mm continuous |
Blendendurchmesser aperture diameter | Ø0,5mm, Ø11mm Ø0.5mm, Ø11mm |
Messbereich measuring range | 1 – 500 Photonenanzahl/ms 1 – 500 photon counts/ms |
Messgenauigkeit measurement accuracy | ±5% (bei 80 Photonenanzahl/ms) ±5% (bei 80 photon counts/ms) |
Beobachtungslichtquelle observation light source | Halogenlampe (12V, 30W) Halogen lamp (12V, 30W) |
Für Flaremeter for Flaremeter | Laserdiode (640nm, 35μ W) Laser diode (640nm, 35μ W) |
Licht Detektor Photomultiplier (PMT) Light Detector Photomultiplier (PMT) | |
Drucker Printer | Thermischer Liniendrucker mit automatischer Papier-schneide-Funktion Thermal line printer with automatic paper cutting function |
Druckerpapier printer paper | 58mm breit 58mm width |
Monitor Monitor | 4,3"inch" TFT Farb LCD Monitor 4.3inch TFT Color LCD Monitor |
Schnittstelle Interface | RS-232C RS-232C |
Abmessungen/Gewicht Dimensions/Weight | 600 (B) x 460(L) x 560 (H)mm / 25kg 600 (W) x 460(L) x 560 (H)mm / 25kg |
Stromanschluss power supply | AC100 - 230V, 50 / 60Hz AC100 - 230V, 50 / 60Hz |
Stromverbrauch power consumption | 60VA / 80VA (maximum) 60VA / 80VA (maximum) |