URK-900F is used to determine the initial, objective refractive values for a patient's eyes in the workflow of refraction and to determine the optical prescriptions for myopia, hyperopia and astigmatism.
3 Dimensional movement mechanism enables to track down a measuring focus of an eye automatically and complete the measurement for eyes by auto shooting automatically. Automatically track pupils and focus to complete the measurement.
URK-900F can determine the best position for precise focusing with the help of high resolution scanner.
The LCD moves freely and can be measured anywhere, LCD Tilting & Swiveling to 270° and provide a monitor reverse function.
It’s equipped with the controller to control 3 Dimension mechanism and auto shooting.
Just by simple handling, users can set the height of measuring point comfortably.
The printer cover can be opened and closed by pushing a button, making it easy to replace a paper. Embedded printer allow print a measurement data without noise. Also, it cuts paper automatically after printing.
You can check abnormal crystalline lenses, cataracts and scratches of corneas to see how healthy patient’s eyes are. With an enforced REF power, you can check SPH, CYL, and Axis that could not be measured in a normal mode.
Ref / Ker / Ref&Ker / Illumination / CLBC / KP Mode, Pupil Size Measurement.
It may measure periphery of cornea both horizontally and vertically from the center of cornea to get curvature and eccentricity value of all points.
It may allow the best fitting for contact lenses.
It provides a reliable measurement by our unique temperature compensation system.
Considering measurement environment by temperature, it may reinforce the accuracy of optometry.
It serves many languages, basically 10 languages. (English, Spanish, Chinese, Japanese, German, French, Italian, Portuguese, Russian and Korean)
Various languages besides the above may be served according to the requirement of customer.
3 Dimensional movement mechanism : enables to track down a measuring focus of an eye automatically and complete the measurement with auto shooting for eyes after focusing automatically.
Auto Focusing System : URK-900F can determine the best position for a precise focusing with the help of high resolution scanner.
8 Inch Touch LCD : URK-900F could examine a patient’s vision accurately with a high speed itself and do auto tracking toward the opposite eye after measuring an eye.
LCD Tilting & Swiveling : The LCD moves freely and can be measured anywhere. LCD Tilting & Swiveling to 270° and provide a monitor reverse function.
Auto Cutting Printer : High speed Printer prints out the final measurement results in 3 seconds.
Peripheral Keratometry Measurement : URK-900F provides peripheral keratometry measurement data that can be greatly useful for fitting contact lenses.
Various Network : Bluetooth, Ethernet, USB, RS-232, Etc.
Convenient S/W Update : S/W update could be done by USB memory easily and conveniently.
Motorized Chinrest
Mode | K/R Mode, Ref. Mode, Ker. Mode, CLBC Mode | |
Ref. | Vertex Distance (D) | 0, 10, 12, 13.5, 15 mm |
Sphere(SPH) | - 25.00 ~ + 22.00 D (0.12, 0.25 d) | |
Cylinder(CYL) | 0.00 ~ +/- 10.00 D | |
Axis | 1 ~ 180° | |
Cylinder Form | -, +, +/- | |
Pupil Distance | 10 ~ 85 mm | |
Min. Pupil Dia. | ø 2.0 mm | |
Ker. | Radius of Curvature | 5.0 ~ 10.2 mm |
Corneal Power | 33.00 ~ 67.50 D | |
Corneal Astigmatism | 0.00 ~ -15.00 D (0.12/0.25 D) | |
Axis | 1 ~ 180° | |
Corneal Dia. | 2.0 ~ 12.0 mm | |
Other | Internal Printer | Thermal / Auto Cutting |
Power Saving | Nominate by 3, 5, 10 Minutes | |
Display | 8.0 inch 24 bit Colored TFT LCD | |
Power Supply | AC 100 ~ 240 V, 50/60 Hz, 50-70VA | |
Dim./Weight | 285(W) x 522(D) x 438(H) mm / 19 Kg | |
Chinrest Mov. | Max. 65 mm Motorized Operating | |
Data Output | RS-232 Interface, USB (for service) |